If the Shoe Fits… Thank UV, Not Cinderella

This article was originally published in UV+EB Technology,

Q3

2024

By Jim Raymont

The proverb “if the shoe fits” first appeared as “if the cap fits” in a poem by Daniel Defoe in 1705. The expression evolved by the late 18th century and eventually became associated with the glass slipper in the fairy tale, “Cinderella.”

Cinderella is a fairy tale with a happy, predictable ending. This column is a real-world tale and is best described as a work-in-progress by the customer, with real-world values and images that are not pretty. The ending of this story still is being written.

 

Images 1 and 2: Examples of shoes produced with UV
Images 1 and 2: Examples of shoes produced with UV
Images 3 and 4:The front and back of an EIT UviCure Plus II are shown. These illustrate “normal” use after several years.
Images 5 and 6:These show the front and back of an EIT UviCure Plus II after just a few months of use. The images show how excessive heat and improper handling have changed the appearance of the unit in a short amount of time.

Shoe Production and UV

 

UV is used to produce the soles of shoes, primarily in Brazil, India and Vietnam. The customer produces 120,000 pairs of soles per day, which equals 5.6 million pairs per month and 43 million pairs per year (Images 1 and 2). The soles in this facility are produced on 17 different UV lines.

The customer has maintained a “we have always done it that way” approach. (Refer to the Q2 2024 UV+EB Technology column, “The Seven Dirty Words,” for more information.) The customer has a policy to measure each of its 17 lines once every two hours. Two different radiometers are used, which equates to each radiometer taking over 3,000 measurements per month.

The facility has significant staff turnover, and our guidelines for proper care of our radiometers are not always followed. Units sent for service/calibration often show excessive wear and exposure to heat of both the cases and optics (Images 3-6).

The changes are not just to the appearance of the instrument housing. The optics window, which is like a lens on a camera or cell phone, also has been damaged by the customer through improper cleaning (Images 7-10).

As a result of improper optics cleaning and the types of damage shown in Images 7-10, customer units frequently are received outside of the expected +/- 10% performance tolerance. Units received read up to 55% high due to customer-induced damage to the optics.

Customer data for the EIT UV-A (320-390 nm) band is shown in Table 1 for units both recently calibrated and units approximately 60 days after calibration. The readings on Serial numbers 17593 and 19498 have increased tremendously because of customer abuse.

Other hidden issues observed by our authorized service and calibration facility in Brazil include the following:

  • Broken battery clips because of the instrument being physically dropped
  • Damaged displays because of the unit being used upside down with the numerical display frequently exposed to the UV sources
  • Damaged electronics and leaking batteries because of repeated exposures to heat far above the recommended temperature limits of the units

 

Table 1: Comparison of units (18657 & 26111) just after calibration with units (17593 & 19498) damaged by improper customer care. Data courtesy of BRCHEMICAL Brazil.
EIT UVA Serial #18657 Serial #26111 Serial #17593 Serial #19498
Run mJ/cm² mW/cm² mJ/cm² mW/cm² mJ/cm² mW/cm² mJ/cm² mW/cm²
#1 289 111 281 97 502 172 456 145
#2 288 107 277 95 505 176 457 150
#3 289 97 270 98 485 180 444 155
Notes Performance on units recently calibrated Performance on units recently calibrated Performance on units with damaged optics, approximately 60 days after calibration Performance on units with damaged optics, approximately 60 days after calibration

 

Image 7: This is an image of a normal EIT optic window. The window is clean, without any damage or customer-deposited material on the optic window.
Image 8: This is an image of a damaged EIT optic window. The textured appearance on the window’s surface is adhesive, which has been UVcured in place.
Image 9: This is an image of a damaged EIT optic window. The customer used a solvent other than isopropyl alcohol (IPA) to clean adhesive off the optic window. The solvent damaged the optic window and created a damaged “ring” on the optics.
Image 10: This is an image of a damaged EIT optic window. When the customer had trouble chemically removing the adhesive with IPA, the UV-cured adhesive was physically removed with an abrasive material, most likely sandpaper. This also removed several layers of material on the EIT optic window.

 


Summary

Most of our customers follow proper use, care and cleaning guidelines and experience much better results, with many EIT instruments lasting more than 15 years with regular service. The instrument performance shown in this column is outside of expectations. The root cause is extreme instrument abuse at the customer facility.

EIT 2.0 and BRCHEMICAL have worked to proactively educate the customer by:

  • Providing multiple education seminars on how to use, handle, clean and care for the instruments,
  • Offering detailed reports when units are received out of tolerance,
  • And asking why the customer feels the need to measure each of its production UV lines every two hours.

The customer does better for periods of time, but frequent staff turnover and lack of a formal training program make consistent performance and a happy fairy-tale ending difficult. Only the customer, with sustained and substantial changes to its internal processes, can change the ending to this tale.

The next time you slip into your shoes, thank UV.

Acknowledgments

Special thanks to BRCHEMICAL (www.brchemical.com) in Brazil for sharing the images and working to break the “we have always done it that way” mentality of this customer. Images 3-10 are courtesy of BRCHEMICAL Brazil.

Parting Thought

If Cinderella’s shoe fit perfectly, then why did it fall off?

Jim Raymont
Director of Sales
EIT 2.0 LLC